2545 retention time shifts with analytical flow rate - WKB82031

Article number: 82031

SYMPTOMS

  • Peak retention time shifts

ENVIRONMENT

  • 2525/2545 Binary Gradient Module
  • 2535/2545/2555 Quaternary Gradient Module

CAUSE

The accuracy of online solvent mixing at analytical flow rates (<1.0 mL/min) is reduced due to large pump head volumes.

FIX or WORKAROUND

Install the following mixer between the pump and the injector.

Mixer (700011657)

clipboard_e9813b0f0aa5e0a126b5b4b139901a0e5.png

ADDITIONAL INFORMATION

The mixer can remain in the flow path for preparative and analytical methods.

id82031, 2535, 2545, 2545Q, 2555