2545 retention time shifts with analytical flow rate - WKB82031
Article number: 82031
SYMPTOMS
- Peak retention time shifts
ENVIRONMENT
- 2525/2545 Binary Gradient Module
- 2535/2545/2555 Quaternary Gradient Module
CAUSE
The accuracy of online solvent mixing at analytical flow rates (<1.0 mL/min) is reduced due to large pump head volumes.
ADDITIONAL INFORMATION
The mixer can remain in the flow path for preparative and analytical methods.

