Syringe plunger overload error on 2767 / 3767 Injector/Collector - WKB7929
Article number: 7929
SYMPTOMS
- Makes 1 - 2one or two injections and then error message appears
ENVIRONMENT
- MassLynx
- 2767 Sample Manager
- 3767 Sample Manager
CAUSE
The sample is viscous, so the syringe speed is too fast.
FIX or WORKAROUND
Lower the syringe aspiration/draw rate:
- Open the inlet method.
- Click Autosampler > Injection tab.
- Lower the Aspiration Speed (%)
Note: Default is 20. - Select File > Save.
If it's not solved by lower the syringe speed, a cleaning of the inject port is needed.
ADDITIONAL INFORMATION
Confirm that the syringe size is correct. In the inlet method, select Waters 2767 > Configuration, and then confirm that the syringe size listed matches the true size.
id7929, 2767, 2767SFC, MLYNX, MLYNXV41, SUPMM