Baseline wanders on UPC2 PDA - WKB69066

Article number: 69066

SYMPTOMS

  • Baseline wanders seen on UPC2 PDA
  • Baseline Before Repair.png
  • Passes leak tests 
  • Passes ABPR health test
  • PDA energy is good

ENVIRONMENT

  • ACQUITY UPC2 PDA Detector

CAUSE

CO2 purity is too low. Tank is only 99.0% pure.

FIX or WORKAROUND

Exchange the CO2 source for CO2 of high enough purity (99.97% minimum).

ADDITIONAL INFORMATION

Refer to page 93 of the ACQUITY UPC2 System Guide (715004521) for more information on system specifications, including CO2 purity.