Baseline wanders on UPC2 PDA - WKB69066
Article number: 69066
SYMPTOMS
- Baseline wanders seen on UPC2 PDA
- Passes leak tests
- Passes ABPR health test
- PDA energy is good
ENVIRONMENT
- ACQUITY UPC2 PDA Detector
CAUSE
CO2 purity is too low. Tank is only 99.0% pure.
FIX or WORKAROUND
Exchange the CO2 source for CO2 of high enough purity (99.97% minimum).
ADDITIONAL INFORMATION
Refer to page 93 of the ACQUITY UPC2 System Guide (715004521) for more information on system specifications, including CO2 purity.