Skip to main content
Waters

Leak test fails on 2545 QGM - WKB32719

Article number: 32719

SYMPTOMS

  • The static leak test fails for both left and right pump heads

ENVIRONMENT

  • MassLynx
  • Autopurification

CAUSE

The QGM mixer is leaking.

FIX or WORKAROUND

Replace the QGM mixer (700004436).

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.