"Energy level too low" error on UPC2 PDA detector - WKB30359
Article number: 30359
SYMPTOMS
- UPC2 PDA fails with error message " Energy level Too Low"
- Error is not cleared by power cycling PDA
- Error is not cleared by flushing flow cell with solvent
- Lamp hours are normal
ENVIRONMENT
- MassLynx
- ACQUITY UPC2 System
- ACQUITY UPC2PDA Detector
- ACQUITY UPC2Convergence Manager
CAUSE
The configured detectors within inlet configuration are in the wrong order
FIX or WORKAROUND
-
The correct order of detectors in the MassLynx Inlet Configuration Wizard is as follows:
Waters ACQUITY CM
ACQUITY Convergence Manager
Waters ACQUITY PDA
ADDITIONAL INFORMATION
This detector order dependency is due to the sequential order in which MassLynx sets up and monitors the instruments before setting up the next module. If the PDA is configured AFTER the column manager but BEFORE the convergence manager, then the PDA setup (exposure calculation/energy check) will occur just after the flow has been started, while the system is still pressurizing. During this time, the PDA will experience "frothing" and initiate the low lamp energy failure.
