Software shows rpm is high but the pump pistons are not moving and pressure is not building - WKB25670
Article number: 25670
SYMPTOMS
- RPM readback in software is very high
- Pump pistons not moving
- Pump does not build pressure
- Pump starvation error
ENVIRONMENT
- P-200 Pump
- SFE Bio-Botanical Extraction System
CAUSE
Timing belt is worn
FIX or WORKAROUND
- Replace the timing belt on the pump—700006540 (Belt, HTD Timing).
ADDITIONAL INFORMATION