Sample loop leaks on 2767/3767 Sample Manager - WKB253488

Article number: 253488

SYMPTOMS

  • Sample loop leaks

ENVIRONMENT

  • 2767 Sample Manager
  • 3767 Sample Manager

CAUSE

Sample loop is worn.

FIX or WORKAROUND

Replace the sample loop (see article wkb242722 for part numbers).