How to teach needle positions after "Needle positions have not been taught" message appears on WFMA - WKB246610
Article number: 246610
OBJECTIVE or GOAL
Teach needle positions after "Needle positions have not been taught" message appears.
ENVIRONMENT
- Waters Fraction Manager - Analytical
PROCEDURE
- Select Console > Fraction Manager-Analytical > Configure > Bed.
- Run Bed reset.
- Select the appropriate plate in the Bed Configuration dialog box, as shown in the picture below.
- Click Teach and follow the instructions on the screen to teach the Above Vessel and In Vessel positions.
Notes:
- When teaching the Above Vessel position, set the needle height low enough to avoid spraying into other vials.
- Set the needle height low to prevent spraying into other vials, and set the needle height high to ensure safe transfer between vial positions.
- Set a height that allows safe transfer between vial positions.
ADDITIONAL INFORMATION