Skip to main content
Waters

No peaks observed in Prep mode on AutoPurification system - WKB241739

Article number: 241739

SYMPTOMS

  • No peaks observed in Prep mode on AutoPurification system 
  • 515 pump shows low system pressure
  • Solvent leaks from splitter box

ENVIRONMENT

  • AutoPurification
  • System Fluidics Organizer
  • 515 HPLC Pump

CAUSE

The splitter delay coil was damaged due to excessively high flow rates.

FIX or WORKAROUND

Replace the splitter coil. 

ADDITIONAL INFORMATION

clipboard_e5e8282006d31685a2fa1e0a9b328b752.png

Note: 700012238 ASSY, CAP TUBING, 19.7 LONG. Is used in both 205000438 Splitter, Flow, 5000:1, 8-30mL  and  205000437 Splitter, Flow, 1000:1, 8-30mL. 

Not able to find a solution? Click here to request help.