No peaks observed in Prep mode on AutoPurification system - WKB241739
Article number: 241739
SYMPTOMS
- No peaks observed in Prep mode on AutoPurification system
- 515 pump shows low system pressure
- Solvent leaks from splitter box
ENVIRONMENT
- AutoPurification
- System Fluidics Organizer
- 515 HPLC Pump
CAUSE
The splitter delay coil was damaged due to excessively high flow rates.
FIX or WORKAROUND
Replace the splitter coil.
ADDITIONAL INFORMATION
Note: 700012238 ASSY, CAP TUBING, 19.7 LONG. Is used in both 205000438 Splitter, Flow, 5000:1, 8-30mL and 205000437 Splitter, Flow, 1000:1, 8-30mL.
id241739, 515, SFO