Syringe plunger stops moving down after a few seconds when priming the 2767 Sample Manager - WKB22280
Article number: 22280
SYMPTOMS
- Syringe plunger starts to move down but stops after a few seconds
- Invalid command error message
ENVIRONMENT
- 2767 Sample Manager
- MassLynx
CAUSE
A clog in the pathway creates resistance when the syringe plunger moves down.
FIX or WORKAROUND
- Disconnect T7 at the valve.
- Put down paper towels.
- Select Waters 2767 > Prime system in the inlet editor.
If the system primes properly and solvent comes out through the valve, T7 is blocked. - Replace the T7 tubing (700002529 - Kit, Tubing T7).
ADDITIONAL INFORMATION