Syringe plunger stops moving down after a few seconds when priming the 2767 Sample Manager - WKB22280

Article number: 22280

SYMPTOMS

  • Syringe plunger starts to move down but stops after a few seconds
  • Invalid command error message

ENVIRONMENT

  • 2767 Sample Manager
  • MassLynx

CAUSE

A clog in the pathway creates resistance when the syringe plunger moves down.

FIX or WORKAROUND

  1. Disconnect T7 at the valve.
  2. Put down paper towels.
  3. Select Waters 2767 > Prime system in the inlet editor.
    If the system primes properly and solvent comes out through the valve, T7 is blocked.
  4. Replace the T7 tubing (700002529 - Kit, Tubing T7).

ADDITIONAL INFORMATION

 