Error with 2767: Invalid plate position, Check plate parameters within bed layout - WKB204996
Article number: 204996
SYMPTOMS
- The following error appears when opening inlet editor: Invalid plate position, Check plate parameters within bed layout
- The 2767 continues to cycle between right-hand side of system and wash station
ENVIRONMENT
- 2767
- MassLynx
CAUSE
Corrupt bed layout in MassLynx
FIX or WORKAROUND
- Inspect all sensors on the 2767 for signs of obstruction.
- Deactivate collection control and delete the fl4 file.
- Deconfigure all systems in the inlet editor.
- Delete all bed layouts from MassLynx.
- Allow the 2767 to initialize.
- Recreate bed layouts and confirm all reference positions.
- Set up collector control.
ADDITIONAL INFORMATION