Leaks from the top of the Fluid Delivery Module (FDM) - WKB202505
Article number: 202505
SYMPTOMS
- Leaks from the top of the Fluid Delivery Module (FDM)
ENVIRONMENT
- SFC Prep 15
- SFC Prep 30
- Investigator
- MV-10
CAUSE
Loose fitting (see attached)
This fitting may become loose little by little.



FIX or WORKAROUND
Regularly (once a year), retighten the fitting by hand.