Pump stops between each stage of the extraction method for a Bio-Botanical Extraction System - WKB12402
Article number: 12402
SYMPTOMS
- Pump stops between each stage of method
- Pump resumes flow rate after a few seconds
ENVIRONMENT
- Bio-Botanical Extraction System (BBES)
- Chromscope 1.6
CAUSE
This is expected behavior.
FIX or WORKAROUND
No resolution is required.
ADDITIONAL INFORMATION
The pumps for the BBE system stop momentarily between stages of the method to allow for the movement of valves. If the valves were to change position amid normal flow rates, the pressure would spike, risking failure of a rupture disk.
id12402, SFEBBC2