Skip to main content
Waters

"Z-axis lost steps" error on 2767 Sample Manager - WKB112027

Article number: 112027

SYMPTOMS

  • "Z-axis lost steps" error on 2767 Sample Manager during injection

ENVIRONMENT

  • 2767 Sample Manager
  • MassLynx

CAUSE

The probe cable is loose or damaged.

If Loose the cable can snag on system tubing and become damaged

FIX or WORKAROUND

  1. Replace the probe cable (part 700001928) if damaged.
  2. Secure the cable with a cable tie.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.

  • Was this article helpful?