Random communication errors - WKB111940

Article number: 111940

SYMPTOMS

  • Random communication errors
  • Loss of communication to various modules
  • Valves move to incorrect position

ENVIRONMENT

  • Autopurification system
  • MassLynx

CAUSE

Corrupted method or project

FIX or WORKAROUND

  1. Create a new project using the default project as a template.
  2. Recreate all methods from scratch.

ADDITIONAL INFORMATION