Peaks do not elute on Prep 350 SFC - WKB109577
Article number: 109577
SYMPTOMS
- Peaks not eluting
- Poor baseline
- Replacing the inlet and outlet CV does not help
ENVIRONMENT
- SFC 350
- P 200
CAUSE
A wrong flow rate of 8 mL/min is being used on the co-solvent pump.
FIX or WORKAROUND
The recommended flow rate for the P200 Co-Sol pump is 20-200 mL/min and 30-350 mL/min for the 350 CO2 pump.
ADDITIONAL INFORMATION
Refer to "Operation of the Prep 350 SFC System with ChromScope".
id109577,