Peaks do not elute on Prep 350 SFC - WKB109577

Article number: 109577

SYMPTOMS

  • Peaks not eluting
  • Poor baseline 
  • Replacing the inlet and outlet CV does not help 

ENVIRONMENT

  • SFC 350 
  • P 200

CAUSE

 A wrong flow rate of 8 mL/min is being used on the co-solvent pump.

FIX or WORKAROUND

The recommended flow rate for the P200 Co-Sol pump is 20-200 mL/min and 30-350 mL/min for the 350 CO2 pump.

ADDITIONAL INFORMATION

Refer to "Operation of the Prep 350 SFC System with ChromScope".