With 2767 Sample Manager, no peaks detected on analytical injections - WKB108920
Article number: 108920
SYMPTOMS
- No peaks seen on UV or MS traces when running analytical injections
- Prep injections are good
ENVIRONMENT
- 2767 Sample Manager
- AutoPurification
CAUSE
Leak on the rotor/stator of the left-hand analytical valve
FIX or WORKAROUND
Rebuild the valve, replacing the rotor and stator using part 201000214.
ADDITIONAL INFORMATION
id108920, 2767, 2767SFC