What is the column connection back lock ferrule design for the three-piece, finger-tight APH fitting? - WKB50981

Article number: 50981

ENVIRONMENT

  • Active Preheater
  • ACQUITY Arc
  • ACQUITY UPLC
  • ACQUITY UPLC I-Class
  • ACQUITY UPLC H-Class PLUS
  • ACQUITY UPLC H-Class PLUS Bio
  • ACQUITY UPLC I-Class PLUS
  • ACQUITY UPLC H-Class
  • ACQUITY UPLC H-Class Bio

ANSWER

The back lock ferrule design for the three-piece APH fitting is 700010009 (10/pk) or 700011767 (2/pk)

ADDITIONAL INFORMATION

For additional information, see Install 3-Piece Finger-Tight Fitting with APH.

The ferrules for these products are not interchangeable. The OD/ID of the compression fitting for Arc and ACQUITY PLUS is different than older ACQUITY products. Older ACQUITY products will not accept the captive ferrule design 700010009. That will only fit in the newer compression screws.

Video Reference: Install 3-Piece Finger-Tight Fitting with APH - http://site-436448.bcvp0rtal.com/detail/videos/how-to-series/video/6033348301001/install-3-piece-finger-tight-fitting-with-aph?autoStart=true