"Seal force sensor h/w fault" with ACQUITY SM-FTN - WKB49431
Article number: 49431
SYMPTOMS
- "Seal force sensor h/w fault" (numbers vary from 6.0 to 10.0)
- SM-FTN fails to initialize
- Cannot reset communications
ENVIRONMENT
- ACQUITY APC Sample Manager - FTN
- ACQUITY Arc
- ACQUITY Arc Premier
- ACQUITY Arc Sample Manager FTN-R
- ACQUITY UPLC H-Class
- ACQUITY UPLC H-Class Bio
- ACQUITY UPLC H-Class PLUS
- ACQUITY UPLC H-Class PLUS Bio
- ACQUITY UPLC I-Class PLUS
- ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)
CAUSE
The spring inside the wash station is stuck or possible salt deposit causing it to be stuck .
FIX or WORKAROUND
- Remove wash station and disassemble and clean if needed.
- Ensure that the spring in the wash station is positioned around the three prongs.
- Run characterize needle seal after the repair. Consider also running the needle seal readiness test
ADDITIONAL INFORMATION
If the inside of the wash station or spring is damaged or corroded consider replacing the entire assembly part number (700005226 Hclass or 700009226 IClass+ & IClass)