Article number: 28983

SYMPTOMS

  • The source heater does not heat above 54 degrees
  • The desolvation heater reaches the set temperature
  • A full System reboot was performed

ENVIRONMENT

  • TQ Detector
  • SQ Detector

CAUSE

Faulty cartridge heater

FIX or WORKAROUND

  1. Place the instrument in Standby, vent the system, and power-off the Detector
  2. Replace the ion block cartridge heater, part number 700003050

ADDITIONAL INFORMATION