Alliance: "A request for instrument configuration has failed" - WKB228386

Article number: 228386

SYMPTOMS

  • Cannot establish a connection with Alliance
  • Alliance visible in DHCP
  • Scanning for instrument in Node properties shows the Alliance status as "Yes"
  • Pinging the IP address of the 2695 is successful—no lost data packet
  • On the Run screen, a message at the bottom states Instrument failure for 2690/5
  • Message center error messages include "A communication failure has occurred" and "A request for instrument configuration has failed"

ENVIRONMENT

  • Alliance HPLC System

CAUSE

Bad Ethernet cable

FIX or WORKAROUND

Replace the unshielded Ethernet cable with a shielded cable (441000456 - 5 ft. cable).

ADDITIONAL INFORMATION