Solvent found below the bottom-left corner of instrument - WKB99721

Article number: 99721

SYMPTOMS

  • Solvent found below the bottom-left corner of instrument

ENVIRONMENT

  • SYNAPT G2
  • SYNAPT G2 MS/HDMS
  • SYNAPT G2-S MS/HDMS
  • SYNAPT G2-Si MS/HDMS
  • SYNAPT G2-S
  • SYNAPT G2-Si
  • SYNAPT XS
  • Xevo TQ-S
  • Xevo TQ-MS

CAUSE

Broken drain elbow

FIX or WORKAROUND

Replace the elbow (700004865) and secure the hose to waste. The elbow must be ordered through Waters service.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.