Excessive hissing noise comes from the QDa exhaust pinch valve assembly - WKB9969
Article number: 9969
SYMPTOMS
- Excessive hissing noise comes from the QDa internal exhaust pinch valve assembly (KAD model)
- Instrument operates normally, no error messages or loss of performance
- Hissing noise is continuous during system operation
- QDa passes the source pressure test if the leak is small, but if the leak is large, the test fails
- High nitrogen consumption noted
ENVIRONMENT
- QDa versions with internal pinch valve assembly
CAUSE
O-ring is fitted incorrectly in the pinch valve assembly, creating a minor nitrogen leak.
The four screws of the pinch valve assembly are unevenly tightened. 
New Design since March 2024

FIX or WORKAROUND
- Remove the pinch valve assembly and re-seat the O-ring (as shown) such that the curved profile faces outward.
- Ensure that the four screws that hold the pinch valve assembly in place are evenly tightened.
- If damaged, replace Pinch Valve (Internal) Part Number 700011173
ADDITIONAL INFORMATION
The leaking nitrogen is not from the source exhaust but from the nitrogen flow used to close the pinch valve; it does not represent a hazard to users.
See related article: Loud noise coming from QDa pinch valve assembly - WKB75987
