Source pressure test leak warning on a Xevo Instrument - WKB9958
Article number: 9958
SYMPTOMS
- Source pressure test leak warning message
- Source pressure test failed message
ENVIRONMENT
- Xevo QTof
- Xevo TQ
- Xevo TQ-S
- Xevo G2 QTof/Tof
- Xevo G2-S QTof/Tof
- Xevo G2-XS QTof/Tof
- Xevo G3 QTof
- Xevo TQ-S micro
- Xevo TQ-S cronos
- Xevo TQ-XS
- Xevo TQ-Absolute
CAUSE
Loose or poorly fitted exhaust tubing between instrument and trap bottle
Loose or poorly fitted fitting (e.g. 4 mm Legris PushFit elbow below the isolation valve)
Trap bottle lid needs to be tightened more
FIX or WORKAROUND
- Inspect the Source exhaust trap bottle during the source pressure test and search for location of potential nitrogen leak (e.g. hissing noise).
- Check that all fittings are correctly fitted to the trap bottle (see the "additional information" section here below).
- Remove and reinstall the different tubing.
- Rerun the source pressure test:
Open and close the source door.
or
Run the source pressure test via the software in use.
ADDITIONAL INFORMATION
Old design of source exhaust trap bottle:

New design of source exhaust trap bottle (shipping from 2022 with Xevo TQ Absolute and Xevo G3 QTof):