Analyzer vacuum failure for Xevo TQ MS after rebooting the instrument for a communication failure - WKB98741
Article number: 98741
SYMPTOMS
- Communication error for Xevo TQ MS in the ACQUITY MS Console
- Hardware reboot of the system
- Analyzer vacuum failure error for the Xevo TQ MS in System Status after the hardware reboot
ENVIRONMENT
- Xevo TQ MS
- Xevo TQD
CAUSE
During the hardware reboot of the Xevo TQ MS, the mass spectrometer was powered-off.
FIX or WORKAROUND
- Open MS Tune from the Instrument Shortcut in MassLynx.
- From the Vacuum menu, select PUMP.
- On the Diagnostics tab of the Tune page, monitor the turbo pump speed.
- When the turbo pump speed reaches 100% for both turbo pumps, the system is ready for use.
ADDITIONAL INFORMATION
