How to calculate the minimum interscan delay times for SIR Voltage experiments carried out on the AutoSpec - WKB96691
Article number: 96691
OBJECTIVE or GOAL
How to calculate the minimum interscan delay times for SIR Voltage experiments carried out on the AutoSpec.
ENVIRONMENT
- AutoSpec Premier
- AutoSpec Ultima
- MassLynx
PROCEDURE
- The SIR Voltage experiment has a mass value, channel time, and delay (interscan) time. The delay time occurs before the channel time.
- The minimum delay time is calculate as 1% change in mass x 1 msec + 5 msec.
- In the example above, the mass 334 goes to mass 317 = 17Da mass change. 17/317 = 5.4% change in mass. Multiplying this by 1 msec = 5.4 msec + 5 msec = 10.4 msec. I recommend that the value be rounded up to the nearest 5 msec, so the calculated delay time should be 15 msec.
ADDITIONAL INFORMATION
- If the minimum delay time is not calculated correctly, the first mass can fall out of relative intensity limits with respect to its M+2 isotope for EPA dioxin methods. Lowest mass ions tend to be due to the polychlor furan compounds.
- Do not assume that canned SIR Voltage experiments that come with MassLynx have the correct calculated delay times.
- Channel times can be altered sometimes to as low as 10 msec. Note that there will be a √2 decrease in signal/noise as a result.
- See WKB95321 on how an SIR Voltage experiment is carried out on the AutoSpec.
