With a Tof MS, how do you quantify with TIC? - WKB94908

Article number: 94908

ENVIRONMENT

  • Xevo G2-XS QTof
  • SYNAPT G2-Si

ANSWER

Quantify with the total value of 50-1000 Da spectra.

When performing the Tof MS quantitation, use the chromatogram extracted at m/z xxx for the target compound ion.


For a chromatogram that combines all spectra in a wide range:
Unless the following conditions are met, it is not generally possible to obtain good linearity, so it is not recommended.

  • Background ions (ions other than the target compound) are always present with a constant intensity
  • The types of background ions are always constant
  • No ions are intense enough to saturate in the background ions

ADDITIONAL INFORMATION