Xevo TQ-XS Sensitivity gradually decreases - WKB93682
Article number: 93682
SYMPTOMS
- Xevo TQ-XS Sensitivity gradually decreases.
- As the time between analysis increases, the sensitivity increases, and it declines in continuous analysis. (MS operation is always ON.)
ENVIRONMENT
- Xevo TQ-XS
- UniSpray
CAUSE
Dirty MS divert valve.
FIX or WORKAROUND
- Confirm diagnosis by bypassing diverter valve:If that restores stable sensitivity, replace Diverter Valve (700004188)
ADDITIONAL INFORMATION