Xevo TQ-XS Sensitivity gradually decreases - WKB93682

Article number: 93682

SYMPTOMS

  • Xevo TQ-XS Sensitivity gradually decreases.
  • As the time between analysis increases, the sensitivity increases, and it declines in continuous analysis. (MS operation is always ON.)

ENVIRONMENT

  • Xevo TQ-XS
  • UniSpray

CAUSE

Dirty MS divert valve.

FIX or WORKAROUND

  1. Confirm diagnosis by bypassing diverter valve:If that restores stable sensitivity, replace Diverter Valve (700004188)

ADDITIONAL INFORMATION

id93682, XEVOG2XSQT, XEVOTQXS, XEVTQXSIVD