DC Bias fails during instrument setup on Vion - WKB93496

Article number: 93496

SYMPTOMS

  • Instrument setup fails at DC Bias section
  • Instrument electronics were rebooted

ENVIRONMENT

  • Vion IMS QTof
  • UNIFI Workstation

CAUSE

Unknown

FIX or WORKAROUND

  1. Take instrument system offline.
  2. In Device Management section of UNIFI, go to the Workstation tab, right-click and select Reset Device.
  3. Bring the instrument system back online.
  4. Confirm that instrument setup can be completed.

ADDITIONAL INFORMATION

 

id93496, SUPUNIFI, UNIFISVR, UNIFISW18, UNIFOPT, UNIFQLIC, UNIFSW17, UNIFSW18, UNIFSW19, UNIFWGLIC, UNIFWKLIC, VION