Low MS sensitivity on QTof mass spectrometers - WKB91752
Article number: 91752
SYMPTOMS
- MS sensitivity is around three orders of magnitude lower than expected for leucine enkephaline (1E3 instead of 1E6) and other analytes.
- Occurs in both positive and negative resolution modes (customer is not using any other modes).
- Detector setup fails.
ENVIRONMENT
- SYNAPT G2-Si
- SYNAPT G2-S
- SYNAPT G2
- Xevo G2-XS Qtof
- Xevo G2-S Qtof
- Vion IMS QTof
CAUSE
Customer uses tune file created for troubleshooting a low-sensitivity problem on high m/z ions. In this tune file, the quad profile is set to a manual profile in both positive and negative resolution mode.
FIX or WORKAROUND
Set quad profile to auto profile or load a tune file with an auto quad profile.
ADDITIONAL INFORMATION
id91752, MALDIG2SI, SYNG2HD, SYNG2HDMAL, SYNG2MS, SYNG2MSMAL, SYNG2SHD, SYNG2SHDML, SYNG2SIHD, SYNG2SIMS, SYNG2SMS, SYNG2SMSML, XEVOG2SQTF, XEVOG2STOF, XEVOG2XSTF