Article number: 91571

SYMPTOMS

  • Detector Setup and Detector Check are missing from the console

ENVIRONMENT

  • Xevo G2-XS
  • Xevo G3 QTof

CAUSE

The Instrument Console was left in ASAP mode.

FIX or WORKAROUND

  1. Open up the source enclosure. The source then detects that an ESI source is present.
  2. Click ESI when prompted.
  3. Navigate to the console and change the Ionization mode from ASAP to ESI.
  4. Detector Setup now appears in the Console.