Low response with APCI probe - WKB90985
Article number: 90985
SYMPTOMS
- Low response from instrument
ENVIRONMENT
- APCI TQD TQS
CAUSE
Blockage in the APCI probe capillary—on close inspection, liquid is coming not from the capillary tip but from between the sleeve and the capillary.
FIX or WORKAROUND
Rebuild the probe with a new capillary.
ADDITIONAL INFORMATION
