Xevo TQ-S source exhaust warning - WKB89906

Article number: 89906

SYMPTOMS

  • Source exhaust high pressure warning seen on tune page
  • Unable to go into Operate with warning

ENVIRONMENT

  • Xevo TQ-S
  • Xevo TQ-S micro
  • Xevo TQD
  • Xevo TQ-XS

CAUSE

Faulty nitrogen supply

FIX or WORKAROUND

  1. Inspect all tubing to and from the rear of the instrument and the trap bottle for kinks or blockages.
  2. Swap the trap bottle and test to see if the source pressure test now completes.
  3. Change nitrogen for a known working supply if thought to be faulty.

ADDITIONAL INFORMATION