Turbo controller fault on SQD2, Xevo TQD, Xevo TQ-S or Xevo TQ-S micro - WKB8866
Article number: 8866
SYMPTOMS
- Error message: "Turbo controller fault"
ENVIRONMENT
- SQD2
- TQD
- Xevo TQ-S micro
- Xevo TQ-S
- MassLynx
CAUSE
Leak on the source after cleaning sample cone.
FIX or WORKAROUND
Reseat or Replace sample cone seal (700004321; O-ring, size 10).
ADDITIONAL INFORMATION
id8866, A-SQ, MLYNX, MLYNXV41, SQD2, SUPMM, TQD, TQDIVD, XEVOTQD, XEVOTQDIVD, XEVOTQS, XEVOTQSIVD, XEVOTQSMIC, XEVTQSMIVD