Cone voltage slider is unavailable (dimmed) on the Tune page - WKB88575
Article number: 88575
SYMPTOMS
- Cone voltage slider is unavailable/dimmed and cannot be adjusted
- All other tuning parameters are available
ENVIRONMENT
- Xevo TQ-XS
- Xevo TQ-S
- Xevo TQ MS
- ACQUITY QDa Mass Detector
CAUSE
Cone voltage ramp is active.
FIX or WORKAROUND
- On the Tune page, either clear the cone voltage ramp shortcut button or click Ramps and Use Cone Voltage Ramp to deactivate.
- Verify that the cone voltage slider can be used as expected.