SYNAPT/SYNAPT G2 will not pump down - WKB87265

Article number: 87265

SYMPTOMS

The scroll pump starts when the pump command is given, but the backing pressure does not reduce.

ENVIRONMENT

SYNAPT G2

CAUSE

The Speedivalves on the manifold top of the scroll pump are closed.

FIX or WORKAROUND

Open the Speedivalves and pump down again.

ADDITIONAL INFORMATION