Nominal mass shifts over time - WKB84196
Article number: 84196
SYMPTOMS
- Nominal masses shift with laboratory temperature
ENVIRONMENT
- All Tof instruments
- Time-of-flight technology
CAUSE
New equipment added in the laboratory is affecting room temperature stability and the mass measurement of the Tof instrument.
FIX or WORKAROUND
Contact the building facility staff to address the room temperature fluctuation within the laboratory.
ADDITIONAL INFORMATION
id84196, Delta, GCT, GCTPREMIER, LCT, LCTPREMIER, LCTPREMXE, MALDIG2SI, QTOF2, QTOFAPIUS, QTOFGLBL, QTOFMICRO, QTOF-PREM, QTOFULTAPI, RDA, SYNAPT, SYNAPTMS, SYNAPTXS, SYNG2HD, SYNG2HDMAL, SYNG2MS, SYNG2MSMAL, SYNG2SHD, SYNG2SHDML, SYNG2SIHD, SYNG2SIMS, SYNG2SMS, SYNG2SMSML, VION, XEVOG2QTOF, XEVOG2SQTF, XEVOG2STOF, XEVOG2TOF, XEVOG2XSTF, XEVOQTOF