ACQUITY TQD randomly vents - WKB83192
Article number: 83192
SYMPTOMS
- The ACQUITY TQD vents with no errors
- Rotary pump has stopped pumping
- Unable to restart pump immediately after crash vent
ENVIRONMENT
- ACQUITY TQD
CAUSE
The rotary pump is overheating, causing it to switch off.
FIX or WORKAROUND
- Allow the rotary pump to cool down to cold.
- Restart the pumping sequence from the tune page.
ADDITIONAL INFORMATION
The rotary pump in question for this article example was a Leybold SV40 BFIC model V3010. Its maximum operating temperature is 40°.
If the pump is in a noise box, ensure that all fans are spinning and that the noise box can regulate temperature without overheating.