Fluidics stops on the MS Console for the Xevo G2-XS - WKB82881
Article number: 82881
SYMPTOMS
- System had run out of nitrogen, but after resolving that issue the MS console is red with a message that the fluidics stopped for the Xevo
- Sample fluidics shows a message to purge to recover
- Reference fluidics shows idle
ENVIRONMENT
- MassLynx
- Xevo G2-XS
- Xevo TQ MS
- Xevo TQ-S
- Xevo TQ-XS
- Xevo TQD
- Xevo TQ-S micro
- Xevo TQ-S cronos
CAUSE
Purge is not selected on the Tune page Fluidics tab.
FIX or WORKAROUND
- Open the mass spectrometer's tune page.
- Go to the Fluidics tab and select the sample fluidics section.
- Click the button to Purge.
- Wait for sample fluidics to be in Idle state, and then test.