How to prepare 0.5 mM sodium hydroxide solution in 90:10 IPA/water from powder or pellets - WKB81422
Article number: 81422
OBJECTIVE or GOAL
Prepare 0.5 mM sodium hydroxide solution in 90:10 IPA/water from powder or pellets (used in making calibration solution 0.5 mM sodium formate solution in 90:10 2-propanol/water)
ENVIRONMENT
- Preparing calibration solution
- IPA, isopropanol, 2-propanol or propan-2-ol
- Xevo QTof
- Xevo G2 QTof/Tof
- Xevo G2-S QTof/Tof
- Xevo G2-XS QTof/Tof
- SYNAPT
- SYNAPT G2-S
- SYNAPT G2-Si
PROCEDURE
- Dissolve 200 mg of NaOH pellets in 1 L water and label "5 mM NaOH in water".
- Transfer 10 mL of the 5 mM NaOH solution to a 100-mL volumetric flask.
- Fill to the mark with 2-propanol (IPA).
- Label the flask "0.5 mM sodium hydroxide solution in 90:10 IPA/water".
ADDITIONAL INFORMATION
SIGMA, > or equal to 98% pellets anhydrous
Acros organics sodium hydroxide micro pearls (99% purity)
Use only UPLC/MS-grade solvents