No lockmass signal seen - WKB81141

Article number: 81141

SYMPTOMS

  • No lockmass signal is seen from the reference sprayer
  • The isolation valve is open
  • Signal is seen with analyte sprayer

ENVIRONMENT

  • Xevo G2-S
  • Xevo G2-XS
  • Xevo G3 QTof
  • MassLynx
  • UNIFI
  • SELECT SERIES Cyclic IMS

CAUSE

Most likely a clog or air in fluidics.

FIX or WORKAROUND

Purge the reference fluidics several times. If still no lockmass signal rebuild reference sprayer PN: 700004674

ADDITIONAL INFORMATION

 

id81141, XEVOG2SQTF, XEVOG2STOF, XEVOG2XSTF, XEVOG3QTOF

 