No lockmass signal seen - WKB81141
Article number: 81141
SYMPTOMS
- No lockmass signal is seen from the reference sprayer
- The isolation valve is open
- Signal is seen with analyte sprayer
ENVIRONMENT
- Xevo G2-S
- Xevo G2-XS
- Xevo G3 QTof
- MassLynx
- UNIFI
- SELECT SERIES Cyclic IMS
- SYNAPT G2-Si
CAUSE
Most likely a clog or air in fluidics.
FIX or WORKAROUND
Purge the reference fluidics several times. If still no lockmass signal rebuild reference sprayer PN: 700004674
ADDITIONAL INFORMATION