Fluidics bottles are overflowing on Xevo TQ-S micro - WKB81095
Article number: 81095
SYMPTOMS
- Solution is overflowing the MS reservoirs
- The fluidics bottles are filling up
ENVIRONMENT
- Xevo TQ-S micro
CAUSE
The divert valve stator is leaking.
FIX or WORKAROUND
Replace the rotor using the diverter rotor/stator kit (700009465).
ADDITIONAL INFORMATION
id81095, XEVOTQSMIC