Source pressure test leak trip for QDa - WKB80261
Article number: 80261
SYMPTOMS
- Source pressure leak trip
ENVIRONMENT
- ACQUITY QDa Mass Detector
CAUSE
Faulty check valve assembly on the bottle trap
FIX or WORKAROUND
- Clean the check valve manifold.
- If cleaning does not help, replace the bottle waste trap assembly (700009635).
ADDITIONAL INFORMATION
id80261, checkvalve, QDA