Source pressure test leak trip for QDa - WKB80261

Article number: 80261

SYMPTOMS

  • Source pressure leak trip

ENVIRONMENT

  • ACQUITY QDa Mass Detector

CAUSE

Faulty check valve assembly on the bottle trap

FIX or WORKAROUND

  1. Clean the check valve manifold.
  2. If cleaning does not help, replace the bottle waste trap assembly (700009635).

ADDITIONAL INFORMATION

 