Skip to main content
Waters

Low sensitivity on some compounds on a Xevo TQ-S - WKB78904

Article number: 78904

SYMPTOMS

Low sensitivity on some compounds on a Xevo TQ-S.

ENVIRONMENT

Xevo TQ-S

CAUSE

Incorrect source offset voltage.

FIX or WORKAROUND

Infuse sample and monitor the signal on the tune page and optimize the source offset voltage to give maximum sensitivity for the compound.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.