SYMPTOMS

  • "QDa Turbo Power Too low" error

ENVIRONMENT

  • ACQUITY QDa Mass Detector

CAUSE

Poor vacuum

FIX or WORKAROUND

Tighten the vacuum clamps between the QDa and the roughing pump.