Console error: Nitrogen gas failure - WKB77409

Article number: 77409

SYMPTOMS

  • Error seen in console: "Nitrogen gas failure"

ENVIRONMENT

  • Quattro Premier
  • Quattro Premier XE
  • 3100 Mass Detector
  • SQ Detector
  • SQ Detector 2
  • TQ Detector
  • ACQUITY TQD
  • Xevo TQ MS
  • Xevo TQD
  • Xevo TQ-S
  • Xevo TQ-S cronos
  • Xevo TQ-S micro
  • Xevo TQ-XS
     

CAUSE

Unstable nitrogen supply, or the instrument draws from a finite supply (such as a dewar) that runs out of N2.

FIX or WORKAROUND

  1. Ensure that the nitrogen supply is at the correct pressure (98 +/- 4 psi).
  2. In the MS Tune page, click the Gas pull-down menu and choose Reset Nitrogen Failure.
  3. Verify that the error message has cleared.