Console error: Nitrogen gas failure - WKB77409
Article number: 77409
SYMPTOMS
- Error seen in console: "Nitrogen gas failure"
ENVIRONMENT
- Quattro Premier
- Quattro Premier XE
- 3100 Mass Detector
- SQ Detector
- SQ Detector 2
- TQ Detector
- ACQUITY TQD
- Xevo TQ MS
- Xevo TQD
- Xevo TQ-S
- Xevo TQ-S cronos
- Xevo TQ-S micro
- Xevo TQ-XS
CAUSE
Unstable nitrogen supply, or the instrument draws from a finite supply (such as a dewar) that runs out of N2.
FIX or WORKAROUND
- Ensure that the nitrogen supply is at the correct pressure (98 +/- 4 psi).
- In the MS Tune page, click the Gas pull-down menu and choose Reset Nitrogen Failure.
- Verify that the error message has cleared.
ADDITIONAL INFORMATION
id77409, XEVOTQ, XEVOTQA, XEVOTQIVD