"Turbo controller fault" on ACQUITY SQD - WKB77316
Article number: 77316
SYMPTOMS
- SQD vents and gives the error "Turbo controller fault" in the instrument console
- The instrument pumps down but fails to reach 100% turbo speed or maintain turbo speed
ENVIRONMENT
- SQ Detector
- Xevo TQD
CAUSE
No oil in rotary pump
FIX or WORKAROUND
- Fill the rotary pump with fresh oil.
- Pump instrument via the tune page.
- Ensure that turbo speed reaches 100%.
ADDITIONAL INFORMATION
The error message "Turbo controller fault" does not necessarily indicate a fault with the turbo controller or turbo pump; it is simply a message from the controller reporting that a reading (e.g., turbo speed, turbo temperature, power supply voltage) is outside of a specified range. The most common problem resulting in this message is a turbo speed that is out of specification due to insufficient backing to the turbo pump.