What are the nitrogen gas requirements for the ACQUITY QDa Mass Detector? - WKB76811
Article number: 76811
ENVIRONMENT
- ACQUITY QDa Mass Detector
ANSWER
The detector requires a supply of dry, oil-free nitrogen with a purity of at least 95%.
The nitrogen must be regulated at 6.5 to 7 bar (94 to 102 psi) outlet pressure, using a two-stage gas regulator with an appropriate outlet range—for example, 0 to 11 bar (0 to 160 psi).
Typical nitrogen usage is 1300 L/h (at atmospheric pressure).