Xevo G2-XS does not stay pumped down - WKB76728
Article number: 76728
SYMPTOMS
- Instrument will pump down with the isolation valve shut
- When the isolation valve is opened, the instrument loses vacuum
ENVIRONMENT
- Xevo QTof
- Xevo G2 QTof
- Xevo G2 Tof
- Xevo G2-S
- Xevo G2-S QTof
- Xevo G2-S Tof
- Xevo G2-XS
- Xevo G2-XS QTof
- Xevo G2-XS Tof
CAUSE
The sample cone O-ring is damaged or dirty.
FIX or WORKAROUND
Clean or replace the sample cone O-ring.
ADDITIONAL INFORMATION
700004321 - O-Ring, 10 x 1.5mm, fluoroelastomer (1/pk)
id76728, XEVOG2QTOF, XEVOG2SQTF, XEVOG2STOF, XEVOG2TOF, XEVOG2XSTF, XEVOQTOF