Skip to main content
Waters

Xevo G2-XS does not stay pumped down - WKB76728

Article number: 76728

SYMPTOMS

  • Instrument will pump down with the isolation valve shut
  • When the isolation valve is opened, the instrument loses vacuum

ENVIRONMENT

  • Xevo QTof
  • Xevo G2 QTof
  • Xevo G2 Tof
  • Xevo G2-S
  • Xevo G2-S QTof
  • Xevo G2-S Tof
  • Xevo G2-XS
  • Xevo G2-XS QTof
  • Xevo G2-XS Tof

CAUSE

The sample cone O-ring is damaged or dirty.

FIX or WORKAROUND

Clean or replace the sample cone O-ring.

ADDITIONAL INFORMATION

700004321 - O-Ring, 10 x 1.5mm, fluoroelastomer (1/pk)

id76728, XEVOG2QTOF, XEVOG2SQTF, XEVOG2STOF, XEVOG2TOF, XEVOG2XSTF, XEVOQTOF

Not able to find a solution? Click here to request help.