Poor injection reproducibility on HDX2 - WKB76474
Article number: 76474
SYMPTOMS
- Poor reproducibility of injection
ENVIRONMENT
- HDX-2
- Chronos
- LEAP PAL-3
CAUSE
A droplet of sample remained on the tip of the needle and was deposited on the internal side of the labeling vial septum.
FIX or WORKAROUND
Increase the "Injection Speed PREP [uL/s]" to 20 (default is 5).
ADDITIONAL INFORMATION
id76474, LEAPPAL