Skip to main content
Waters

Poor injection reproducibility on HDX2 - WKB76474

Article number: 76474

SYMPTOMS

  • Poor reproducibility of injection 

ENVIRONMENT

  • HDX-2
  • Chronos
  • LEAP PAL-3

CAUSE

A droplet of sample remained on the tip of the needle and was deposited on the internal side of the labeling vial septum.

FIX or WORKAROUND

Increase the "Injection Speed PREP [uL/s]" to 20 (default is 5).

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.